Analysis Default Conditions
The name of the Stage position file used in an automatic
analysis is defined here.
Sample Description and Analytical Conditions
Each element has its own experimental
conditions (Accelerating Voltage and/or Beam Current). Elements can be
measured by either EDS or WDS. Elements not measured can have their compositions
Known, calculated by Difference or by Stoichiometry.
Example of a Peak Integration analysis with Quantitative
Interference
Below V Ka is measured by Peak Integration.
The same element (V Ka) interferes with the Kb X-ray
line of Ti.
The menu defined the Interferences
for each element measured (maximum five interferences per element). Note
that each element which interferes is associated with a standard, this
will allow the correction program to calculate the real intensity of the
element measured.
|